Medical Device Alert

07 May 2008

Device:
Disposable pen torch manufactured by Merlin Medical Limited and distributed by Williams Medical Supplies.
Model Number: W2137
NHS Supply Chain Reference (England Only): FFE066.
 
Problem:
Incorrect bulb placement due to a manufacturing fault which can cause the bulb to explode upon illumination or be damaged during normal use.
 
Action:
  • Identify and isolate all pen torches with the bulb protruding more than 1.5mm beyond the tip of the torch.
  • Contact the supplier to arrange for any affected product to be replaced with updated stock.
  • Ensure users are aware of the updated information card which states that the pen torch should be tested away from the patient before use.
For further information and contacts please see the MHRA website.